Catalog Number 80239 |
Device Problems
Retraction Problem (1536); Stretched (1601); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a recanalization procedure in the right femoral artery via the left femoral access, when the physician retracted the device out of the sheath, the tip of the device and the inner helix allegedly stretched apart more than the normal.It was further reported that the device was allegedly difficult to be retrieved through the sheath.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4: (expiration date: 11/2025).H11: section a: through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a recanalization procedure in the right femoral artery via the left femoral access, when the physician retracted the device out of the sheath, the tip of the device and the inner helix allegedly stretched apart more than the normal.It was further reported that the device was allegedly difficult to be retrieved through the sheath.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the catheter was returned for evaluation.The catheter was physically investigated.During physical investigation, the tip of the catheter was stuck with bodily material and test guidewire passed until the tip of the catheter.After running in the water slightly lower than nominal aspiration level was achieved.Mechanical jam was observed during the investigation.Therefore, the investigation cannot be confirmed for the reported stretch and retraction problem.A clear root cause could not be identified but a blockage of the catheter represents a known inherent risk.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|