• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE INFUSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 2426-0500
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd alaris pump module smartsite infusion set separated the following information was received by the initial reporter with the following verbatim: product description: bd alaris pump infusion set origin of the issue: product failure//design detail: the connector that is between the top y-site connector and the safety clamp in the pump segment of the tubing is coming apart, should be glued.
 
Manufacturer Narrative
Sample inv and dhr the customer reported separation and returned one used sample of material 2426-0500 and lot 23103166.The set was visually examined for defects and abnormalities.The separation in silicon segment at upper fitment of the pump segment was verified by visual inspection.Under a microscope, the ring retainer indent on the silicon was unable to be found.The ring retainer creates the binding for this connection, there is no solvent or glue.The manufacturing location found several root causes for the lack of ring retainer, including the retainer fixture for silicon tubing was dirty, incorrect set up, operator omits placing the ring, sensors did not detect missing ring.Actions taken by the plant are to improve the detection of the component by the sensors, and also to improve the cleaning process of the silicon.Device history record review for model 2426-0500 lot number 23103166 was performed.The search showed that a total of (b)(6) units in 1 lot number was built on 23oct2023.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS PUMP MODULE SMARTSITE INFUSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18587053
MDR Text Key333843245
Report Number9616066-2024-00074
Device Sequence Number1
Product Code FPA
UDI-Device Identifier37613203021006
UDI-Public(01)37613203021006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2426-0500
Device Lot Number23103166
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/05/2024
Initial Date FDA Received01/26/2024
Supplement Dates Manufacturer Received02/28/2024
Supplement Dates FDA Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-