MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Arrest (1762)
Event Date 09/12/2023
Event Type  Injury  
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found no abnormalities.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported while using the insufflator the patient experienced cardiac arrest for 5 to 6 seconds and the therapeutic laparoscopic right hemicolectomy procedure was canceled.It was further clarified, the cardiac arrest lasted 25 seconds and the set pressure was 10mmhg at that time.The patient was transferred to kyoto katsura hospital and underwent "lipectomy".The surgery was completed without any problems.The patient had a history of atrial fibrillation, however the doctor thought it unlikely a direct relation to the cardiac arrest.It was reported, the cardiac arrest happened when head-down."it was assumed that the cardiac arrest was caused by compression due to the combined factors of positional change and insufflation.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18587368
MDR Text Key333845572
Report Number3002808148-2024-00799
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/08/2024
Initial Date FDA Received01/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age77 YR
Patient SexFemale
Patient Weight58 KG
-
-