Model Number PM3562
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that the patient presented for a follow-up in the clinic.Upon interrogation, the pacemaker was unable to be interrogated.The pacemaker was explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported event unable to interrogate was not confirmed.Final analysis found that the device received above elective replacement indicator (eri) level.Electrical and mechanical analysis performed indicated normal functionality with normal inductive and rf telemetry.Additionally, visual inspection of the header and connector found no contamination or foreign material that could contribute to the reported event.Longevity assessment was performed, and the device was in normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18592864
MDR Text Key333896211
Report Number2017865-2024-02155
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberPM3562
Device Lot NumberP000151426
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1458QL/86 QUARTET; 2088TC/52 TENDRILLEAD; 2088TC/58 TENDRILLEAD
Patient Outcome(s) Required Intervention;
Patient Age43 YR
Patient SexFemale
-
-