Model Number 011050-10
Device Problems Component Missing (2306); Detachment of Device or Device Component (2907); Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during use after 10 minutes, the loop detached from the side poles to which it is attached.Possible fragments not recovered.No harm to patient, user or third reported.
 
Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Manufacturer Narrative
The product was returned on 2024-03-04.The investigation of the product was completed on 2024-03-20.The device product history records (dhr) have been checked for the available lot number and found to be according to the specification, valid at the time of production.The problem described by the customer can be confirmed.The working electrode is molten down completely - the neutral electrode is partially molten.There is no indication for a material-, manufacturing- or design-related failure.Most likely, the damage has been caused by excessive force applied to the cutting loop during application, led to a break of the cutting electrode and by touching the neutral electrode creating a shortcut which, in the end, melts the electrodes.The root cause of the reported issue can most likely be traced back to a usage-related failure.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODE, BIPOLAR
Type of Device
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLSR, SINGLE USE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key18596855
MDR Text Key333976021
Report Number9610617-2024-00020
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K882270
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number011050-10
Device Catalogue Number011050-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2023
Initial Date FDA Received01/29/2024
Supplement Dates Manufacturer Received03/20/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-