DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number EMAX2PLUS
Device Problems Material Frayed (1262); Vibration (1674)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition that the device exhibited vibration, identified during service and repair, was confirmed.The assignable root cause was determined to be due to component failure from wear.(b)(4).
 
Event Description
It was reported by japan that during service and evaluation, it was determined that the motor device had a worn bearing, produced unexpected noise, exhibited vibration, was loose and had cord damage.It was further determined that the device failed pretest for visual assessment, loctite assessment and noise assessment.It was noted in the service order that the device had hose damage.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2024.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: h6: in addition to the failures included in the initial medwatch report, additional pretest failures have been added.Upon further investigation, it was noted that the device also had illegible labeling etch.This information does not change the assignable root cause of component failure due to wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 
3035526892
MDR Report Key18596954
MDR Text Key334026283
Report Number1045834-2024-00106
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384002055
UDI-Public00845384002055
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/24/2024
Initial Date FDA Received01/29/2024
Supplement Dates Manufacturer Received04/09/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-