MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Power Up (1476); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id 37651 lot# serial#(b)(6) implanted: explanted: product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 37651, serial/lot #: (b)(6) ubd: , udi#: (b)(4)medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the recharger was not powering on, and when they checked the equipment they noticed that the desktop charger (dtc) cord was damaged.The caller mentioned that the patient has 2 rechargers (one for each ins), so they have been using 1 recharger to charge each ins.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 37612 serial#(b)(6) implanted: (b)(6) 2018: product type implantable neurostim ulator product id 37651 serial# (b)(6): product type recharger product id 37761 serial# (b)(6): product type recharger h3.Analysis was performed on [product id 37761 serial# (b)(6) analysis found that the cable assembly was frayed and the connector pin was bent.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|