Model Number MX9505 |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that while preparing to use a pressure sensor to monitor central venous pressure, the monitor has no normal waveform after the pressure sensor is connected to the lead wire.At the same time, cannot adjust it to zero.The same problem occurs after replacing the lead wire multiple times.After replacing it, the display is normal.There was no patient involvement reported.
|
|
Manufacturer Narrative
|
B3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device evaluation: one used, decontaminated device was returned for investigation.Visual inspection of the electrical board found one lead wire were was observed to be not fully soldered.The sample failed the electrical test.The complaint was confirmed.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.This issue will continue to be monitored and further actions taken accordingly.
|
|