This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon of "nozzle is clogged" was confirmed; however, it could not be further identified.There was no information available to determine if reprocessing steps may have been a factor.Therefore, a further cause could not be presumed.The suggested phenomenon of "adhesive around the forceps cover is missing" could not be further presumed from the information obtained in this investigation.As a result of confirming the contents of the instruction manual about shipping products, we found the detailed explanation about reprocessing below: chapter 7 cleaning, disinfection, and sterilization procedures; chapter 8 reprocessing workflow for endoscopes and accessories; chapter 9 reprocessing the endoscope(and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
|