• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KAVO DENTAL GMBH GENTLEPOWER LUX CONTRA ANGLE 25LPR; DENTAL HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH GENTLEPOWER LUX CONTRA ANGLE 25LPR; DENTAL HANDPIECE Back to Search Results
Model Number 25 LPR
Device Problems Overheating of Device (1437); Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Event Description
During a restorative treatment to teeth #28 - #30 the patient jumped in the chair and complained about pain due to heat.The handpiece was very hot, and patient's lip was red on the area where the handpiece was touching the lower lip left side.The area did not blister, hence no injury and no medical care necessary.
 
Manufacturer Narrative
During a visual inspection prior to the repair, it was found that the head was dented in several places around the push button and around the water spray insert.This shows that the instrument received strong external hits, e.G., due to dropping.During the test run the heat up was reproducible.After disassembling it was found that the bearings have been worn out running rough.This caused unusual inner friction and hence heat up of the back cap.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.Caution: burning hazard from hot instrument head or hot instruments cover.If the instrument overheats, burns may arise in the oral area.Never contact soft tissue with the instrument head or instrument cover the following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems.Before each use, the contra-angle handpiece must be checked for external damage.Before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration.Immediately stop using contra-angle handpieces that act unusual.Never press the push button during operation.This also includes lifting the cheek or tongue! to ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used.Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.This service interval depends on the frequency of use and should be adjusted accordingly.Service may only be carried out by kavo-trained repair shops using original kavo replacement parts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLEPOWER LUX CONTRA ANGLE 25LPR
Type of Device
DENTAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, 88400
GM   88400
MDR Report Key18604250
MDR Text Key334177379
Report Number3003637274-2024-00003
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073478
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Dentist
Type of Report Initial
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25 LPR
Device Catalogue Number1.007.3580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2024
Initial Date FDA Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-