• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXFORD TRL BRG W/SLOTS LGE 3MM; BMET GENERAL KNEE INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXFORD TRL BRG W/SLOTS LGE 3MM; BMET GENERAL KNEE INSTRUMENTS Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4) g2 - foreign: canada investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that at the end of the trailing, the trial bearing broke during removal.This event is related to a malfunction that could potentially lead to a serious injury.However, no patient harm or further outcome was reported.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.Product was not returned in its original packaging.Part shows little wear and some damage.The material on the bottom face of the instrument location slot has fractured on both sides.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD TRL BRG W/SLOTS LGE 3MM
Type of Device
BMET GENERAL KNEE INSTRUMENTS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18604790
MDR Text Key334142549
Report Number3002806535-2024-00033
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279346875
UDI-Public(01)05019279346875(11)230720(10)ZB7367177
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-422708
Device Lot NumberZB7367177
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-