It was reported that the patient programmer (pp) showed a 'call your dr.' message and stated that the call your dr.Message cleared away.Then the pp screen went blank and wouldn't power on with 2 sets of new batteries.The patient stated that he turns stimulation up to drive and said when stimulation is turned up he can't talk. since the pp doesn't power on, he wasn't able to adjust stimulation currently; therefore, he was "stuck" at home not being able to drive until the replacement patient programmer arrived.
|
Continuation of d10: product id 37642, serial# (b)(6).Product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 37642, serial/lot #: (b)(4), ubd: 23-may-2014, udi#: (b)(4).H3: analysis of the 37642 patient programmer (ptm) (serial number (b)(6)) revealed corrosion.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|