• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems Device Alarm System (1012); Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2024
Event Type  malfunction  
Manufacturer Narrative
D5.Operator of device is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device was leaking by the "straining tube and keep alarming".No patient or clinical injury was reported.
 
Manufacturer Narrative
Device evaluation: one device was received.Per visual inspection, the enclosure was cracked around the quick connect and the device was leaking from the crack, quick connect was corroded.Per functional testing, the device is leaking from the quick connect where the crack was but could not confirm the customer complaint that the alarm kept alarming.The root cause for the leak was the cracked enclosure.It was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced enclosure, quick connect, reservoir gasket, o-rings.Replaced membrane switch and hl-30 label.Performed preventative maintenance (pm) and calibration.The device passed all functional and delivery tests after repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18607637
MDR Text Key334177849
Report Number2183161-2024-00115
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/10/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/03/2024
Initial Date FDA Received01/30/2024
Supplement Dates Manufacturer Received02/21/2024
Supplement Dates FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-