MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ S CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTBB1D4 |
Device Problem
Pacing Problem (1439)
|
Patient Problems
Bradycardia (1751); Endocarditis (1834); Sepsis (2067); Heart Failure/Congestive Heart Failure (4446)
|
Event Date 12/15/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of d10: 620bg27 heart band medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient was admitted to the hospital due to heart failure with a septic blood infection.The source of the infection was bacterial endocarditis.The sepsis was treated with six weeks of antibiotics.At the time of admission, a dialysis port was placed through the same vein the leads were placed in ¿ the right subclavian vein.The physician believed this damaged the right atrial (ra) lead, causing oversensing and undersensing.Small p waves or noise caused the cardiac resynchronization therapy defibrillator (crt-d) to not properly pace and a loss of crt pacing occurred resulting in bradycardia.This further lead to heart failure, fluid overload, and an irregular heart rhythm resulting in short-long-short episodes that were proarrhythmic.The patient stopped their antiarrhythmic and high blood pressure medications, and their sodium and potassium levels were out of range, thus, making pro-arrythmia more likely.Approximately four weeks after the hospital admission, the ra lead was reprogrammed leading to one hundred percent crt pacing.The device and lead remain in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was further reported that both the atrial and right ventricular (rv) leads had vegetation.The device, the atrial, rv and left ventricular (lv) leads were explanted and replaced one week later due to the endocarditis infection.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|