Catalog Number 72200616
Device Problems Electrical /Electronic Property Problem (1198); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Event Description
It was reported that during set-up, a mdu´s motor was burned out; meaning it overheat.There was a back-up device available and a delay of less than 30 minutes took place.No patient involvement.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
Internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection revealed a discolored cable.Discolored cable is a cosmetic issue that does not affect device functionality and it is not related to the alleged failure.A functional evaluation was performed on the returned device and a handpiece sensor fault error was found.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately.The root cause was associated with an electrical component failure.Factors that could have contributed to the reported event include a failure of the reed switch or damage on the hall board which may contribute to a short circuit.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18607768
MDR Text Key334262791
Report Number1643264-2024-00073
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-