A customer reported that during a patient procedure, using a glidescope core 10-inch monitor, the screen froze several times and then the image went to half-screen causing a delay in patient care.A bflex single-use bronchoscope via glidescope core quickconnect cable and a glidescope spectrum single-use laryngoscope via glidescope core smart cable were being used with the monitor to intubate the patient but the customer reported being unable to isolate the issue to a particular system component.No use of a backup device or harm to the patient was reported.
|
The customer declined the option to have their glidescope core 10-inch monitor returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Upon verathon following up with the customer for further troubleshooting, the customer confirmed being unable to isolate the issue to a particular system component but that their biomedical department would do further troubleshooting to make a decision about sending it to verathon for evaluation.To date, no response has been received from the customer indicating they would like to return their system for evaluation.Review of complaint history for the reported monitor serial number "(b)(4)" did not identify any previous complaints reported to verathon.Trending analysis for glidescope core monitors does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|