Catalog Number 886-42584-05
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2023
Event Type  malfunction  
Event Description
The event involved a transpac® iv monitoring kit, disposable transducer, 03 ml squeeze flush device, macrodrip where it was reported there was leakage on tubing.No report of patient harm.
 
Manufacturer Narrative
The device is not available for investigation however a photo is available for the investigation.
 
Manufacturer Narrative
The reported complaint of leakage was not confirmed.No sample was returned for evaluation.However, two (2) images were provided showing the involved device inside the packaging and the label of the device.A failure mode was not able to identified from the images provided.A lot history was reviewed and no commodities were identified which could lead to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18609345
MDR Text Key334773971
Report Number9617594-2024-00065
Device Sequence Number1
Product Code DRS
UDI-Device Identifier00840619068615
UDI-Public(01)00840619068615(17)260301(10)13596341
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K061573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number886-42584-05
Device Lot Number13596341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-