ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; No Match Back to Search Results
Model Number CDVRA500Q
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Discomfort (2330)
Event Date 01/10/2024
Event Type  Injury  
Event Description
It was reported that the patient presented for follow up.An interrogation of the implantable cardioverter defibrillator revealed inappropriate shock delivered by device.Inappropriate therapy was attributed incorrect diagnosis of supraventricular tachycardia with rapid ventricular response.Programming interventions were made.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18611242
MDR Text Key334138610
Report Number2017865-2024-03376
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model NumberCDVRA500Q
Device Lot NumberP000146366
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/10/2024
Initial Date FDA Received01/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA
Patient Outcome(s) Other;
Patient SexFemale
-
-