Continuation of d10: product id wr9200 lot# serial# (b)(6) implanted: explanted: product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: wr9200, serial/lot #: (b)(6), ubd: , udi#: h3: analysis of the wr9200 recharger (serial number npw019056n) revealed device was unresponsive.Flash sector read/write protection bits have become set.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
It was reported that the patient (pt) stated they just received the replacement recharger one week ago and today it won't turn on the green lights just keep flickering.Pt said the other one was doing the same thing.Patient services (pss) worked with pt to reset the charger, off and on the dock, the dock green light was on the patient confirmed they use the thick cord but the issue was not resolved.Offered and sent replacement request for the recharger and the dock, cord and ac power adaptor due to the same issue happening again.The issue was not resolved.No patient symptoms were reported.
|