• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA ACCESS; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA ACCESS; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD568805908
Device Problems Break (1069); Detachment of Device or Device Component (2907); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  malfunction  
Manufacturer Narrative
Event site telephone: (b)(6).Additional information will be provided following the conclusion of the investigation.H3 other text : device not returned to manufacturer.
 
Event Description
On (b)(6) 2024 getinge became aware of an issue with one of surgical lights - volista access.As it was stated during checks performed on the lamp, it was noticed that spring arm cover has come off.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination or serious injury.
 
Manufacturer Narrative
The correction of b5 describe event and problem, h6 medical device ¿ problem code, h6 investigation findings and h6 investigation conclusions deems required.This is based on the internal evaluation.Previous b5 describe event and problem: on 26th january, 2024 getinge became aware of an issue with one of surgical lights - volista access.As it was stated during checks performed on the lamp, it was noticed that spring arm cover has come off.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination or serious injury.Corrected b5 describe event and problem: on 26th january, 2024 getinge became aware of an issue with one of surgical lights - volista access.As it was stated during checks performed on the lamp, it was noticed that spring arm cover has come off.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination or serious injury.Further information provided by getinge employee indicated that the spring arm cover didn't came off, but was removed and no particles were missing.Based on additional input from getinge employee it was possible to determine that the issue investigated herein is not safety and risk related, as there was no indication of fallen parts, which was initially considered.Therefore, the scenario described in the record is considered as non-reportable.Previous h6 medical device ¿ problem code: mechanical problem|detachment of device or device component||2907 / material integrity problem|break||1069 corrected h6 medical device ¿ problem code: no apparent adverse event|||3189 previous h6 investigation findings: results pending completion of investigation|||3233 corrected h6 investigation findings: no device problem found|||213 previous h6 investigation conclusions: conclusion not yet available||11 corrected h6 investigation conclusions: no problem detected||67 initially provided information was pointing to spring arm cover came off from device.The issue is considered as safety related as any parts falling off into sterile field or during procedure may cause contamination or serious injury.According to additional clarification provided by the getinge technician, the initial information was incorrect.It was determined that the issue investigated herein is not safety and risk related as there was no indication of cover came off- the spring arm cover was removed and no particles were missing.The investigation was performed.The investigated scenarios did not cause risk to human life.The review of the customer product complaints, related to investigated issue in time, shows that there is no regular income.No apparent reason was identified for suggesting to open a capa or evaluation for the need of another action in the market.
 
Event Description
On 26th january, 2024 getinge became aware of an issue with one of surgical lights - volista access.As it was stated during checks performed on the lamp, it was noticed that spring arm cover has come off.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination or serious injury.Further information provided by getinge employee indicated that the spring arm cover didn't came off, but was removed and no particles were missing.Based on additional input from getinge employee it was possible to determine that the issue investigated herein is not safety and risk related, as there was no indication of fallen parts, which was initially considered.Therefore, the scenario described in the record is considered as non-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA ACCESS
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key18612821
MDR Text Key334181232
Report Number9710055-2024-00108
Device Sequence Number1
Product Code FSY
UDI-Device Identifier3700712420505
UDI-Public(01)3700712420505
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568805908
Device Catalogue NumberARD568805908
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-