SYNTHES GMBH RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.240 |
Device Problems
Break (1069); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported from serbia that during an unspecified surgical procedure, a sagittal saw device was used.According to the report, the screw pins/teeth that lock the saw blade in place were broken and missing.There were no delays in the surgical procedure.It was not reported if a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Investigation summary : a visual and functional assessment was performed on the device.The following steps failed: section 50: check the saw blade coupling.During the service evaluation the following failures were identified: device interaction (2+ devices) : cannot secure/lock cutter.Conclusion: failure reported is not confirmed root cause : faulty parts (3210) action :repair, device history review : no manufacturing record evaluation required as no manufacturer or service related issue found.
|
|
Search Alerts/Recalls
|
|
|