Model Number NIMBUS II PAINPRO |
Device Problems
Mechanical Problem (1384); No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/23/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
A review of the device history record has been completed.The pump passed all previous tests.Complaint data was reviewed, there are no previous complaints on this device.Device return requested.This mdr will be reopened and updated in the event the device involved or additional information becomes available.
|
|
Event Description
|
On 01/23/2024, infutronix received a report that a pump stopped infusing due to a system error alert.The infusion cannot resume without causing delay in treatment.Requested device to be returned.
|
|
Manufacturer Narrative
|
Dhr was reviewed, and the pump passed all previous tests.There are no other complaints on this device.The pump was received on (b)(6)2024 and tested on (b)(6)2024.The results are below: the pump's event log was reviewed, and it confirms that there was a system error during an infusion.It took place during a demand bolus delivery.This is seen on line 97 "alarm code: 02 sub code: 24." the subcode indicates motor is fast, motor delay issue.Attempted to resume the infusion that was running on the pump, and it alarmed system error immediately.The reported issue is confirmed.The pump does not meet passing criteria.
|
|