It was reported that corroded batteries damaged the patient programmer's battery compartment.The caller thinks the patient programmer was not powering on as a result of the battery compartment issue, but they could not confirm since they did not have the patient programmer with them.The caller thinks the issue was first noticed at a doctor's appointment in january of 2022, but they could not confirm the event date.The serial number of the patient programmer was uncollectable because the caller did not have it with them.The caller stated that they would call back to provide the serial number of the patient programmer, at which time they would provide the shipping address for the replacement patient programmer.No symptoms were reported.
|
Continuation of d10: product id 37642 lot# serial# (b)(6) implanted: explanted: product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 37642, serial/lot #: (b)(6), ubd: , udi#: (b)(4) h3: analysis of the 37642 programmer (serial number (b)(6)) revealed corrosion.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|