ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 72081-01
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 12/01/2023
Event Type  Injury  
Event Description
The customer reported that the filament of the abbott diabetes care (adc) device remained in their skin.The customer observed that the insertion site was infected and their spouse, who is a nurse removed the filament for treatment.No medication or further treatment was reported.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The reported product is not expected to be returned as reporter indicated the device was discarded.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that freestyle libre sensors continue to be safe, effective, and perform as intended in the field.A tripped trend review was conducted for the reported complaint and freestyle libre sensors, no trends were identified that would indicate any product related issues.The date of event is unknown.The date entered in section b3 is based on the customer's report of "before christmas".The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.In the event that unanticipated product is received, a physical investigation will be performed per adc's established processes and procedures and a follow-up report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.Visual inspection was performed on both sensor patch and sensor tail and no issues were observed.Severed sensor tail was not observed.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If partial product is returned, the case will be re-opened, and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
The customer reported that the filament of the abbott diabetes care (adc) device remained in their skin.The customer observed that the insertion site was infected and their spouse, who is a nurse removed the filament for treatment.No medication or further treatment was reported.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18621612
MDR Text Key334311251
Report Number2954323-2024-03981
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
K223435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number72081-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 01/23/2024
Initial Date FDA Received02/01/2024
Supplement Dates Manufacturer Received02/27/2024
Supplement Dates FDA Received03/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-