• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXALT MODEL D SINGLE-USE DUODENOSCOPE; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXALT MODEL D SINGLE-USE DUODENOSCOPE; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00542421
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Manufacturer Narrative
Block h6 (device codes): problem code a06 captures the reportable event of loss of visualization inside the patient.
 
Event Description
It was reported to boston scientific corporation that an exalt model d scope was used on an endoscopic retrograde cholangiopancreatography (ercp) performed on (b)(6) 2024.About 20 minutes into the procedure, the scope stopped displaying an image.The procedure was successfully completed using another exalt scope.There have been no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block h6 (device codes): problem code a06 captures the reportable event of loss of visualization inside the patient.Block h10: with all the available information boston scientific concludes that the reported event was able to be confirmed, as loss of visualization was replicated during analysis.The returned device was visually inspected; no external defects were identified.The scope was plugged into an exalt model d controller; a live, clear image was displayed.A flush test was performed, and the image remained consistent.However, after connecting the scope a second time, there was no visualization.The reported complaint was confirmed.During the electrical testing, no shorts in the circuit were detected.After disassembling the handle, evidence of residue was found between pins 4 and 5 in the repeater pcb (printed circuit board).After the residue was removed, the scope was re-connected to the controller and a live image was displayed.This investigation is assigned a most probable conclusion code of cause trace to component failure which indicates that the problems were traced to a random component failure with no evidence of a design or manufacturing issue.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event Description
It was reported to boston scientific corporation that an exalt model d scope was used on an endoscopic retrograde cholangiopancreatography (ercp) performed on (b)(6) 2024.About 20 minutes into the procedure, the scope stopped displaying an image.The procedure was successfully completed using another exalt scope.There have been no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXALT MODEL D SINGLE-USE DUODENOSCOPE
Type of Device
DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18621839
MDR Text Key334314126
Report Number3005099803-2024-00267
Device Sequence Number1
Product Code FDT
UDI-Device Identifier08714729993605
UDI-Public08714729993605
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00542421
Device Catalogue Number42421
Device Lot Number0032920760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-