BOSTON SCIENTIFIC CORPORATION INGENIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
|
Back to Search Results |
|
Model Number J177 |
Device Problems
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535); Impedance Problem (2950); Inappropriate or Unexpected Reset (2959)
|
Patient Problem
Asystole (4442)
|
Event Date 01/22/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the patient with this pacemaker was seen in the hospital for a routine follow-up visit.During device interrogation, a reset was triggered due to an increase in impedance, and an asystole of approximately 10 seconds was observed.Boston scientific was contacted and recommendation to replace the device was suggested since the patient is pacemaker dependent.The patient was kept in the hospital.Subsequently, the device was explanted and replaced.There were no additional adverse patient effects reported.The device is expected to return for analysis.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Event Description
|
It was reported that the patient with this pacemaker was seen in the hospital for a routine follow-up visit.During device interrogation, a reset was triggered due to an increase in impedance, and an asystole of approximately 10 seconds was observed.Boston scientific was contacted and recommendation to replace the device was suggested since the patient is pacemaker dependent.The patient was kept in the hospital.Subsequently, the device was explanted and replaced.There were no additional adverse patient effects reported.The device was returned for analysis.
|
|
Search Alerts/Recalls
|
|
|