Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that after the procedure the screwdriver was found to be broken with the tip broken off inside the black set screw.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: updated: b4; b5; g3; h2; h3; h6.Visual examination of the returned product identified the tip of the device had fractured below the threads.There is a wear line on the shaft and on the tip near the fracture site.The complaint is confirmed based on the evaluation of the returned device.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|