Catalog Number 010002749
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2024
Event Type  malfunction  
Event Description
It was reported that after the procedure the screwdriver was found to be broken with the tip broken off inside the black set screw.It was reported that no further information is available.
 
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: updated: b4; b5; g3; h2; h3; h6.Visual examination of the returned product identified the tip of the device had fractured below the threads.There is a wear line on the shaft and on the tip near the fracture site.The complaint is confirmed based on the evaluation of the returned device.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STRAIGHT MOD SCRDRVR 3.5MM
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18622678
MDR Text Key334322438
Report Number0001825034-2024-00251
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304926912
UDI-Public(01)00880304926912(11)170504(10)ZB170202
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number010002749
Device Lot NumberZB170202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-