• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION ZOLL X SERIES CARDIAC MONITOR; MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION ZOLL X SERIES CARDIAC MONITOR; MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM) Back to Search Results
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2024
Event Type  malfunction  
Event Description
The blood pressure monitor read out of service needs repair when turned on at (b)(6) er for transport.Unable to process the blood pressure.The monitor check at time of truck check off had preformed without a issue.Blood pressures taken manually for the transport time and after the transport the monitor was checked and nothing found.The battery was removed and put back in place, this appeared to have fixed the issue and the monitor was left in service.The oncoming crew briefed on the situation and advised that all report and forms have been filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL X SERIES CARDIAC MONITOR
Type of Device
MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
MDR Report Key18623356
MDR Text Key334439758
Report NumberMW5150964
Device Sequence Number1
Product Code DRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/31/2024
Patient Sequence Number1
Patient Age90 YR
Patient SexFemale
Patient Weight52 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-