|
Catalog Number SEPX-8-6-40-135 |
Device Problem
Malposition of Device (2616)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/29/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that an attempt was made to use a protege rx to treat a plaque lesion in the proximal carotid artery - common. degree of tortuosity was moderate. degree of calcification was moderate.Lesion stenosis was 70%.Device was prepped per ifu.Resistance was not encountered when advancing the device.Excessive force was not used. inaccurate delivery occurred.Stent deployed in unintended lesion site.There was sufficient distal landing zone.Thumbscrew/lock-pin was checked for securement prior to procedure.Procedure completed by using another stent.No patient injury reported.
|
|
Manufacturer Narrative
|
Product analysis a portion of the stent was returned to the laboratory it measured as approx.30mm, dried biologics were observed on the returned stent medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|