MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D) Back to Search Results
Model Number G158
Device Problems Signal Artifact/Noise (1036); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Under-Sensing (1661)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 01/19/2024
Event Type  Injury  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) device was in ventricular tachycardia (vt) storm.The patient received inappropriate shocks and anti-tachycardia pacing (atp).There was a signal artifact monitor (sam) episode noted and noise was seen on the right atrial (ra) channel.Upon review, the therapy was given in ventricular tachycardia (vt) due to ventricular was greater than the atrial.The patient was intubated and not well.There was undersensing seen during the atrial fibrillation (af).Technical services (ts) recommended programming options.The patient was in a vt storm and needed to be managed medically to stabilize them.This device remains in service.No additional patient adverse effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18624450
MDR Text Key334337190
Report Number2124215-2024-05674
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public00802526534904
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG158
Device Catalogue NumberG158
Device Lot Number294566
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age83 YR
Patient SexMale
-
-