INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP PROBE 1L M BOLT; ICP MICRSOSENSORS - CERELINK Back to Search Results
Catalog Number 826851
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that when resetting the sensor, the orange light flashes after the user presses the 0 button for 8 seconds; after which the light does not stop on green but on orange.They tried changing the direclink module but encountered the same problem; they changed the sensor and it worked.No patient impact; however, the event led to 30 minutes surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERELINK ICP PROBE 1L M BOLT
Type of Device
ICP MICRSOSENSORS - CERELINK
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18625205
MDR Text Key334532655
Report Number3013886523-2024-00024
Device Sequence Number1
Product Code GWM
Combination Product (y/n)N
PMA/PMN Number
K173192
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number826851
Device Lot Number6806101
Date Manufacturer Received01/15/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-