The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - evaluation found no device deficiencies that would have contributed to the reported complaint as slippage could not be duplicated.When the clamp is properly positioned and put under pressure, it would not move.Unit does have a slight movement in the lock while in the unlocked position; therefore, the disk ratchet and retaining ring have been replaced as a precaution along with general cleaning and maintenance performed.Root cause - the complaint is not confirmed.There is minimal movement in the lock, but it would not cause the unit to slip.Probable root cause of the reported complaint is improper or suboptimal placement of the skull clamp on the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|