MAUDE Adverse Event Report: DEPUY SPINE INC VIPER2 STRAIGHT ROD-480MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER2 STRAIGHT ROD-480MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 186789480
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problems Foreign Body In Patient (2687); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Event Description
Device report from synthes reports an event in japan as follows: it was reported that this was a posterior fusion on (b)(6) 2024.In the surgery, the gripping portion of the rod broke off when the rod in question was removed from the rod inserter.The surgeon determined that because it was not completely separated, it could not be removed.He also determined that the rod migrates within the body from the area.The surgery was completed successfully with no surgical delay.No further information is available.This report is for one (1) viper2 straight rod-480mm this is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D2b: additional product code: kwq, osh, kwp, mnh, mni d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: g1, h4.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part# 186789480.Lot # csnbem.Manufacturing site: medos int.Spine.Supplier: depuy synthes sales, inc.Release to warehouse date: 07 jun 2023.Expiration date: na.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
There has not been a revision or a removal-surgery.There were no adverse consequences that affected the patient.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER2 STRAIGHT ROD-480MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INT. SPINE
chemin-blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key18629995
MDR Text Key334424255
Report Number1526439-2024-00802
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034133522
UDI-Public(01)10705034133522
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number186789480
Device Lot NumberCSNBEM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2024
Initial Date FDA Received02/02/2024
Supplement Dates Manufacturer Received02/13/2024
04/02/2024
Supplement Dates FDA Received02/28/2024
04/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-