• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. SYVA® EMIT ® 2000 TACROLIMUS; ENZYME IMMUNOASSAY, TACROLIMUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. SYVA® EMIT ® 2000 TACROLIMUS; ENZYME IMMUNOASSAY, TACROLIMUS Back to Search Results
Model Number N/A
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Manufacturer Narrative
An outside united states (ous) customer contacted a siemens remote support center (rsc) to report quality control (qc) failures and erroneous patient results with syva® emit® 2000 tacrolimus results that were obtained on viva-e® analyzer.The following maintenance and troubleshooting steps were performed: checked cuvettes, dry block, mixer bearing, belt , valve kits, maintenance diluter¿ replaced cuvette sand dry block, replace wash solution, replace 0.1n hcl,10%naclo n hcl,10%naclo, decontamination of system with micro clean, do checkr&check scv%<1micro clean.Checkr&check scv%<1.Replaced the reagent and repeated tests.The limitations section of the syva® emit® 2000 tacrolimus assay instructions for use (ifu) states: "results of this test should always be interpreted in conjunction with the patient¿s medical history, clinical presentation and other findings." siemens is investigating.
 
Event Description
The customer reported syva® emit ® 2000 tacrolimus results for multiple patient samples on the viva-e® analyzer, that were considered erroneous to repeat testing on the same analyzer.Only results considered correct by the laboratory were reported to the physician(s).Additionally quality control (qc) recovered within range after reprocessing several times.Patient samples were not processed when qc was out of range.Patient results were reported only when qc was in range.There are no known reports of patient intervention or adverse health consequences due to event.
 
Manufacturer Narrative
Siemens filed the initial mdr 2517506-2024-00047 on 02-feb-2024.Additional information 21-mar-2024: siemens performed initial testing and referred the data for further review.Additional information 12-apr-2024: siemens investigated this issue and provided the following conclusion: the expected behavior of the syva® emit® 2000 tacrolimus assay reagent is as follows: information summarized below can be attained in the product instructions for use (ifu), instrument application sheet, and vendor¿s control documentation.The reagent ifu states that the reagents will remain stable after opening for 12 weeks or until the expiration date printed on the label (whichever is sooner), pending that they are stored at 2 ¿ 8°c, upright, and with caps tightly closed.Ifu states to refer to application sheet for complete instructions.Application sheet states that reagents may be stored onboard, capped when not in use for up to 4 weeks or as quality control (qc) results are within acceptable limits.Qc expected ranges are provided by the vendor, with the expectation that individual laboratories will develop their own qc ranges using the expected ranges as a guideline.The syva ifu also provides instruction for the testing laboratory to generate their own qc ranges and states that in the event of a qc failure the testing laboratory is required to follow the laboratory¿s quality procedures.The customer is alleging instability in reagent observed by qc results drifting/failing low and inconsistent clinical sample results.When qc results drift/fail low the customer is required to re-calibrate and to perform additional troubleshooting.Since the atellica ch lot 130034 contains the same reagent used in the syva emit 2000 tacrolimus reagent lot s1 and the viva-e and the atellica ch use the same principle of spectroscopy to measure analyte concentration in calibrator, control, or patient sample, siemens performed testing using the atellica ch reagent lot 130034.The principle of spectroscopy used is based on measuring the absorbance of light at 340 during the reaction between sample, reagent 1, and reagent 2.Siemens' review of the atellica ch data did not confirm the observations reported by the customer.The syva emit 2000 tacrolimus assay reagent has no calibration interval claims and the viva-e application sheet instructs the operator to ¿prepare a calibration curve whenever a new lot of reagents are used or as indicated by control results¿.The product problem identified by the customers is not confirmed.Customer sample handling methods cannot be ruled out as a potential result in adverse assay performance.The syva emit 2000 tacrolimus assay lot s1 is performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.In section h6, the investigation findings and investigation conclusions codes were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYVA® EMIT ® 2000 TACROLIMUS
Type of Device
ENZYME IMMUNOASSAY, TACROLIMUS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive po box 6101
newark DE 19714 6101
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
7818564812
MDR Report Key18630346
MDR Text Key334427793
Report Number2517506-2024-00047
Device Sequence Number1
Product Code MLM
UDI-Device Identifier00842768001666
UDI-Public00842768001666
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K060385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number10445397
Device Lot NumberS1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-