Model Number G158
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Syncope/Fainting (4411)
Event Date 01/04/2024
Event Type  Injury  
Event Description
It was reported that this patient underwent an ablation procedure.During the procedure the patient experienced syncope and it had to be stopped.Upon review of this cardiac resynchronization therapy defibrillator (crt-d) device data, there were stored episodes which coincided with the procedure due to the electromagnetic interference (emi) noise and oversensing which resulted in up to three seconds of pacing inhibition.This patient is pacing dependent.A magnet was confirmed to be used during the procedure.No additional adverse patient effects were reported.Currently, this crt-d remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18630787
MDR Text Key334431786
Report Number2124215-2024-05881
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public00802526534904
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/27/2020
Device Model NumberG158
Device Catalogue NumberG158
Device Lot Number202747
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/23/2024
Initial Date FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age78 YR
Patient SexFemale
-
-