Model Number 1054260
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator's nurse call missing, inlet screws cracked, back enclosure cracked and back bottom handle slight crack.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint of missing nurse call jack was confirmed.The customer does not want any repairs done to the unit.Unit will be returned unrepaired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18631103
MDR Text Key334434521
Report Number2518422-2024-05589
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959015364
UDI-Public00606959015364
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1054260
Device Catalogue Number1054260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2022
Date Manufacturer Received01/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/29/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-