Model Number G151 |
Device Problems
Pacing Problem (1439); Data Problem (3196)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this device electrograms (egm) were flat.Technical services (ts) reviewed noting this device had been programmed to electrocautery protection mode due to a right ventricular (rv) lead issue.Ts noted the egms are flat due to sensing being off in electrocautery protection mode and that tachy therapy had been turned off.Ts noted there were previous inappropriate anti-tachycardia pacing (atp) episodes.This device and rv lead remain in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that this device electrograms (egm) were flat.Technical services (ts) reviewed noting this device had been programmed to electrocautery protection mode due to a right ventricular (rv) lead issue.Ts noted the egms are flat due to sensing being off in electrocautery protection mode and that tachy therapy had been turned off.Ts noted there were previous inappropriate anti-tachycardia pacing (atp) episodes.This device and rv lead remain in service.No adverse patient effects were reported.Additional information was received that this device was explanted.No additional adverse patient effects were reported.
|
|