• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Fluid/Blood Leak (1250); Electrical Overstress (2924)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d5: operator of device are unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that there was exterior leakage and the current value is high.Patient involvement unknown.
 
Manufacturer Narrative
H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Device evaluation: one device was received.Per visual inspection, cracks were observed on the cover of the circulating water tank.Per functional testing, the product's enclosure leakage current value was within the allowable range, and no abnormalities were detected.However, abnormal noises from the circulating water pump and cracks in the tank lid were observed.The complaint was not confirmed.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.At the customer's request, the product was returned to the customer without repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18642598
MDR Text Key334653267
Report Number3012307300-2024-00548
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90JP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/10/2024
Date Manufacturer Received02/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/12/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-