MERIT MEDICAL SYSTEMS MEXICO IMPRESS®.BRAIDED CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 56538MW2/JPB
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/26/2024
Event Type  Injury  
Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow-up report will be submitted once the evaluation is complete.
 
Event Description
The account alleges that during an embolization procedure for internal iliac artery the tip of the catheter detached within the patient.The catheter was inserted from femoral artery and delivered to the internal iliac artery.During the procedure, a guidewire penetrated the side of the tip of the catheter, causing the tip detachment.It was found that the wire went out from the catheter before reaching its tip hole.The physician moved the wire backward and forward within the catheter several times, and it was found that the wire deviated from the catheter every time.Thus, the catheter was attempted to be pulled backward.At that time, the detachment was confirmed.The major portion of the catheter was withdrawn out, and it was found to be broken off at the junction between the black tip and the purple shaft.The vessel had slight calcification as well as tortuosity.The detached portion has been remaining within iliolumbar artery, and it has not been retrieved due to a physician's decision.
 
Manufacturer Narrative
The suspect medical device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key18646318
MDR Text Key334620445
Report Number3011642792-2024-00010
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number56538MW2/JPB
Device Lot NumberI2593042
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-