This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the cracked distal end and corrosion under the lg-lens glue could not be identified.However, it¿s likely the cause of the cracked distal end is related to physical stress by hitting/dropping the distal end, chemical stress from a chemical agent used, or the like.Also, it¿s likely the cause of the corrosion under the lg-lens glue is related to the lg-lens glue peeling due to physical stress by hitting/dropping the distal end, chemical stress from a chemical agent used, or the like; subsequently, a gap was made between the glue and c-body.Then, corrosion was produced by residual water, foreign material, etc.Olympus will continue to monitor field performance for this device.
|