C.R. BARD, INC. (COVINGTON) -1018233 BALLOON BIPOLAR TEMPORARY PACING ELECTRODE CATHETER Back to Search Results
Catalog Number 006173P
Device Problem No Pacing (3268)
Patient Problem Exposure to Body Fluids (1745)
Event Date 12/28/2023
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that there were no signal in temporary pacing electrode catheter.The user has been exposed to harmful substances, blood, or bodily fluids.It was unknown what medical intervention was provided.
 
Manufacturer Narrative
The reported event was inconclusive.The conditions of the sample did not allow further evaluation.Visual evaluation of the returned photo sample noted one opened (without original packaging), used temporary pacing electrode catheter.Visual inspection of the photo sample noted that the reported failure could not be evaluated therefore this investigation is inconclusive.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿open circuit or short circuit¿.However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: precautions: ¿ excessive bending, torquing, or kinking of the electrode catheter may cause damage to the catheter including damage to internal wires.Inspection instructions: visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage.Electrical connections for pacing: 1.Insert adapter pin into standard 2 mm (0.080¿) pin receptacle of equipment (see figure 1).If equipment contains a locking mechanism such as a collet or thumbscrew, tighten down onto adapter.Leave affixed to equipment.Warning: inappropriate electrical connections, e.G., into a wall socket, may pose serious risk of adverse health consequences or death.2.Thread leads of catheter through the adapter eyelet.Connect the negative jack (marked distal) to the negative terminal of the external pulse generator, and the positive jack (unmarked) to the positive terminal of the pulse generator.Correction: d,e h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that there were no signal in temporary pacing electrode catheter.The user has been exposed to harmful substances, blood, or bodily fluids.It was unknown what medical intervention was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLOON BIPOLAR TEMPORARY PACING ELECTRODE CATHETER
Type of Device
TEMPORARY PACING ELECTRODE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18650836
MDR Text Key334656564
Report Number1018233-2024-00520
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K800298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number006173P
Device Lot NumberGFHT0136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2024
Initial Date FDA Received02/06/2024
Supplement Dates Manufacturer Received05/24/2024
Supplement Dates FDA Received05/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-