The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
The reported event was inconclusive.The conditions of the sample did not allow further evaluation.Visual evaluation of the returned photo sample noted one opened (without original packaging), used temporary pacing electrode catheter.Visual inspection of the photo sample noted that the reported failure could not be evaluated therefore this investigation is inconclusive.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿open circuit or short circuit¿.However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: precautions: ¿ excessive bending, torquing, or kinking of the electrode catheter may cause damage to the catheter including damage to internal wires.Inspection instructions: visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage.Electrical connections for pacing: 1.Insert adapter pin into standard 2 mm (0.080¿) pin receptacle of equipment (see figure 1).If equipment contains a locking mechanism such as a collet or thumbscrew, tighten down onto adapter.Leave affixed to equipment.Warning: inappropriate electrical connections, e.G., into a wall socket, may pose serious risk of adverse health consequences or death.2.Thread leads of catheter through the adapter eyelet.Connect the negative jack (marked distal) to the negative terminal of the external pulse generator, and the positive jack (unmarked) to the positive terminal of the pulse generator.Correction: d,e h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|