• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE Back to Search Results
Catalog Number 09043276190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Manufacturer Narrative
The customer¿s e601 module serial number was (b)(6).The e601 module serial number used at the investigation site was b8x5-04.The competitor method was abbott.The ft4 iv reagent lot number used at the investigation site was 724896 with an expiration date of 31-jun-2024.The sample was requested for further investigation.
 
Event Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ft4 iv (ft4 iv) on a cobas e 601 module compared to a competitor method.The sample was submitted for investigation where discrepant ft4 iv results were identified between an e601 module used at the investigation site and the competitor method.Refer to the attached data for the patient results.
 
Manufacturer Narrative
The sample was submitted for investigation.The customer's ft4 results were reproduced.Upon further investigation of the patient sample, no interfering factors were identified.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination, and other findings.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 IV
Type of Device
FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18652890
MDR Text Key336064997
Report Number1823260-2024-00344
Device Sequence Number1
Product Code CEC
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K220456
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09043276190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2024
Initial Date FDA Received02/06/2024
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-