Block e1: (b)(6).Block h6 (device codes): problem code a06 captures the reportable event of loss of visualization inside the patient.Block h10 upon receipt of this scope at our quality assurance laboratory, this device was thoroughly analyzed.An image test was conducted by connecting the umbilicus to a monitor, the monitor showed the image as expected, the image is not lost after the flush test, there were no leaks detected.There were no abnormalities found during the electrical tests.The reported allegation was unable to be confirmed.The device history record (dhr) confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.
|