A visual and functional inspection was performed on the returned device.The reported difficult to flush was not confirmed.The device was prepped to check for any leaks.There were consistent bubbles noted in the syringe upon prepping the device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation was unable to determine a conclusive cause for the reported difficult to flush; however, it is possible the account was referring to the observed leak/splash.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
It was reported that after opening the 7.0mmx19mmx80cm otw omnilink elite stent delivery system (sds), the device was attempted to be prepped, but it was not possible to flush the device as usual.The device was not in a patient.Return device analysis revealed there were consistent bubbles noted in the syringe upon prepping the device.No additional information was provided.
|