• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT Back to Search Results
Model Number 990063-020
Device Problems Signal Artifact/Noise (1036); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the 990063-020 mapping catheter with lot number 227506189 was returned and analyzed.Visual inspection of the loop segment area showed that the loop was kinked and ribbed near the electrode eight.The functional test was performed using a multimeter and the mapping catheter was connected to the test cable.The continuity and impedance measurement between electrodes and the other side of the cable showed the electrocardiogram (ecg) electrode two to pin two was an open circuit.The rest of the channels were normal.Dissection of lemo connector assembly revealed that wire two was moved from soldering inside of the lemo connector.In conclusion, the reported signal noise issue was confirmed through testing.The mapping catheter failed the returned product inspection due to a broken wire observed at the lemo connector and a kink observed at the tip/loop of the pebax tubing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, there was noise on electrodes one to four.The mapping catheter cable was replaced without resolution.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER - 20 MM
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18663668
MDR Text Key335237089
Report Number9617601-2024-00015
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number227506189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2024
Initial Date FDA Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-