Diagnostic/functional testing was performed at the philips authorized repair facility.The bench repair technician (brt) confirmed that speaker has a damaged connector, giving a speaker malfunction and made no sound.A good faith effort (gfe) conducted confirmed that the speaker was not making any sound.Upon opening the speaker ground was detached from the speaker itself and did not make any sound after opening up the unit.The cause of the reported problem was a defective speaker.The reported problem was confirmed.The brt replaced the (453564673831, mx_100/x3 speaker assembly) to resolve the issue.The device has been returned to the customer.
|