The most probable root causes associated with this failure mode are software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The customer reported network error.The reported issue was investigated per doc49410 revision b and attempted to be replicated using similar configurations of samsung galaxy note 10 with android 12 for app version 3.5.1.10363.The reported issue was unable to be replicated and the system functioned as intended.There were no issues identified with the freestyle libre 3 app during replication that would have led to the reported issue.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
An unspecified issue was reported with the adc device in use with a samsung galaxy note 10 with android operating system version 12.As a result, the customer was unable to monitor glucose with sensor readings and experienced symptoms described as vomiting, fatigue, and felt generally unwell and was unable to self-treat, requiring hcp administration of unspecified treatment.No further information was provided.There was no report of death or permanent impairment associated with this event.
|