It was reported that during a ventricular tachycardia (vt) ablation procedure, what appeared to be weird pacing occurred that induced ventricular fibrillation (vf).An external defibrillator was used to convert the vf.Therapy from the cardiac resynchronization therapy defibrillator (crt-d) had been programmed off.Technical services (ts) was contacted, and ts discussed possible explanations such as if the ablation catheter was close to the left ventricular (lv) lead body, the high frequency energy could get capacitively coupled onto the lead conductors and transmit through the lead to the tip or pulse generator.The crt-d system remains in service.No additional adverse patient effects were reported.
|
It was reported that during a ventricular tachycardia (vt) ablation procedure, what appeared to be weird pacing occurred that induced ventricular fibrillation (vf).An external defibrillator was used to convert the vf.Therapy from the cardiac resynchronization therapy defibrillator (crt-d) had been programmed off.Technical services (ts) was contacted, and ts discussed possible explanations such as if the ablation catheter was close to the left ventricular (lv) lead body, the high frequency energy could get capacitively coupled onto the lead conductors and transmit through the lead to the tip or pulse generator.The crt-d system remains in service.No additional adverse patient effects were reported.
|