Product event summary: the data files and the afapro28 balloon catheter with lot number 20144 were returned and analyzed.The patient file showed seven applications were performed using the first balloon catheter identified as afapro28 balloon catheter of lot 20144.Additionally, the patient file showed ten applications were performed using the second balloon catheter identified as afapro28 balloon catheter of lot 20144.The patient file showed a system notice 50032(the safety system has detected a compromised outer vacuum) during ablation at application number seven.External visual inspection of the balloon segment showed blood/fluid inside the balloon.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for seven applications on the reported event date.During functional testing, the console terminated the application and triggered a system notice 50005(that the safety system detected fluid in the catheter and stopped the injection).During pressure testing of the balloon segment, a double-balloon breach condition was identified.Dissection did not show any signs of lifted thermocouple wires or leak detection wires that could have caused the breach.No anomalies were identified during inspection and/or pressure testing of the shaft segment.The pull wire, shaft, and guide wire lumen were intact.In conclusion, the balloon catheter failed the returned product inspection due to a double balloon breach.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|