Model Number N/A
Device Problems Battery Problem (2885); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2024
Event Type  malfunction  
Event Description
It was reported that outside of surgery the pulsavac battery case was bulging with dust spilling into the container.The container had visible black dust inside.There was no patient involvement with no report of harm or delay.Due diligence is in process.No additional information has been received.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: singapore evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d9, g3, g6, h2, h3, h4, h6, h10.Visual examination of the returned product identified the battery pack was busted open and there was black debris from the battery expulsion throughout the tyvek tray.Batteries were in the correct orientation inside the pack.Due to the damage from the expulsion, it could not be determined if there was damage present to the wiring before the expulsion.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18666885
MDR Text Key334842160
Report Number0001526350-2024-00208
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375154
UDI-Public(01)00889024375154(17)260814(10)72198809
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00515047500
Device Lot Number72198809
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received04/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-